You are buying used RADIONICS:MODEL RFG-4A RESEARCH RF LESION GENERATOR SYSTEM This generator was removed from working laboratory, it is cleaned well and been tested by a professional technician, guaranteed to be fully operational.
Radionic RFG-4A Radiofrequency(RF) lession Generator is indicated for use in procedures to create radiofrequency lesions for the treatment of pain, or for lesioning nerve tissue for functional neurosurgical procedures. Radionic RFG4 Radiofrequency Generator is used with separately approved Radiofrequency Probes.Has thermocouple of 30°C to 100°C and timer divison of 1DIV per 5sec with 1rpm.
DESCRIPTIONS
-Thermal RF (carrier frequency)
Auto Temp: Off, 37–95°C
Auto Timer: 5 sec – 30 min
Auto Stepped: Start, Step, Final Temps and Times
Power: 0–50 Watts
-Pulsed RF (carrier frequency)
Auto Temperature: Off, 37–95°C
Auto Timer: 5 sec – 30 min
Voltage: 0-50 Volts Peak
Rates: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 Hz
Widths/Durations: 2, 3, ... , 30 msec
E-dose: Control Voltage or Pulse Width
-Stimulation (Biphasic Square Pulses)
Rates: 2, 5, 50, 75, 100, 150, 180, 200 Hz
Widths/Durations: 0.1, 0.5, 1, 2, 3 msec
Voltage Control: 0-3 Volts
Current Control: 0-10 mA
Impedance Monitor: 30–3000 Ω
Temperature Monitor: 0–110 °C
Additional information:The LG lesion generator series consists of state-of-the-art, portable RF lesion generators for all common lesion techniques (continuous RF and pulsed RF with multiple control modes) in pain treatment and functional neurosurgery. All critical application parameters are permanently monitored by the device, ensuring maximum safety and ease of use during thermocoagulation.
The temperature control of the RF lesion generators allows the performance of controllable thermocoagulations. The volume of the lesion is defined by the diameter of the cannula, the size of the active tip and by the target temperature, making it reproducible. A temperature sensor built into the tip of the TC electrode continuously measures the temperature occurring in the tissue. The RF lesion generators can perform continuous RF outputs in a temperature- or power-controlled mode or via a preset temperature profile. They can also be used in pulsed RF mode with temperature, pulse duration or voltage selection.
Integrated electrical stimulation and impedance measurement
In addition to thermocoagulation, the localisation and differentiation of sensory and motor nerves is possible. Electrical stimulation is used for even more precise electrode positioning. With the RF lesion generators, thermocoagulation, electrical stimulation and impedance measurement can be performed quickly, precisely and safely.
Thermocoagulation procedures for reduction of chronic pain
Interventional pain treatment uses minimally invasive interventions aimed at permanently reducing chronic pain by selectively interrupting or altering the conduction of pain in the affected nerve pathways. This is done either using targeted medication injections or by interrupting pain conduction using minimally invasive neurodestructive procedures - thermocoagulation or cryolesion.
Chronic pain e.g. in facet joint syndrome, can be treated extremely precise, effective and safe by thermocoagulation if the indication is set properly.